Alco-Sensor IV Accuracy Test Instructions

 

1.  Attach the regulator to the dry gas container

 

·       If you are using a new tank, replace the O-ring that is on the regulator’s grooves with the O-ring attached to the new tank; follow with next instructions

·       If using an existing tank, screw the regulator into the top of the tank; be careful not to over tighten

 

2.  Purge the tank for 2-3 seconds prior to the first test

·       Press the top button on the regulator to let gas pass through for 2-3 seconds

 

3.  Insert a new mouthpiece into the Alco-Sensor IV

 

4.  Connect the Mouthpiece to the Regulator

 

·       Make sure that the Alco-Sensor IV displays TEST before moving on

·       Make an airtight seal between the regulator’s delivery tube and the intake tube of the mouthpiece

 

5.  Testing the accuracy

 

·       Depress the button on the top of the regulator for 7 seconds

·       On the 5th second depress the MANUAL button on the Alco-Sensor IV to take a test

·       Stop delivery of the gas on the 7th second

 

6.  Carefully remove the mouthpiece from the regulator

7.  Observe the 3-digit reading

8.  If the 3-digit reading is not acceptable, go to directions for CALIBRATION